6. Interface
The following settings can be made:
Settings
ACTIVE
BAUD
DATA-BITS
STOP-BITS
PARITY-BITS
6.4
Commands for serial interface
All interface commands are preceded by a colon and completed with the carriage
return (CR).
Command and parameters must be separated by a space.
By adding a question mark to the appropriate action the parameter can be read
rather than changed.
Received commands are acknowledged with [OK], unavailable or false commands
with ERROR.
Command
Auto mode
:saaz <0 | 1>
:acy <1 ... 100>
:asd <1 ... 100>
:asu <1 ... 100>
:ate <0 ... 10000>
:ath <1 ... 10000>
:atp <1 ... 10000>
WIKA operating instructions, model CPC2000
Meaning
Selection of the serial interface
The following can be selected:
USB
RS-232
Selection of the baud rate
The following can be selected:
1200
4800
2400
9600
Defined value (8)
Defined value (1)
Defined value (N)
Function
Auto zero before each cycle
(phase 4)
Cycles to be carried out
Steps down
Steps Up
Delay time at the end point
(phase 13)
Hold time (phase 10 and
16)
Pause time (phase 18)
14400
28800
19200
38400
Response
0 --> Off
1 --> On
1 ... 100 --> number of cycles
1 ... 100 --> number of steps to reach
the full scale
1 ... 100 --> number of steps to reach
the full scale
1 ... 10000 --> time in seconds
1 ... 10000 --> time in seconds
1 ... 10000 --> time in seconds
EN
56000
57600
31