7. Faults
7. Faults
Personnel: skilled personnel
If faults cannot be eliminated by means of the listed measures, take the
instrument out of operation immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter
9.2 "Return".
For contact details, see chapter 1 "General information" or the back page
of the operating instructions.
In the event of any faults, first check whether the instrument is mounted correctly,
mechanically and electrically.
Faults
Instrument is not working,
the display is dark
Set pressure has not been
reached, the pump is
running continuously
Instrument smells
when used or becomes
unusually hot when
charging
Instrument smells during
storage or gets unusually
hot
WIKA operating instructions, model CPC2000
Causes
No voltage supply
System is not fully sealed; too
large hose diameter or hose
length
Defective rechargeable battery
Defective rechargeable battery
Measures
Check whether the power
supply unit is connected
properly.
Rechargeable battery is empty
and must first be recharged
Slide hoses on correctly;
eliminate any leaks
Hose diameter max.
6 mm (see chapter
10 "Specifications")
Take the instrument out of
service immediately and
contact the manufacturer
Contact the manufacturer
EN
35