6. Interface
Command
:ssl <d | e>
Menu settings
:smm <a | c | m >
:ssw <1 ... 100>
:swm <a | c | m | z l
v | s>
:szc <0.1>
:szi <1 ... 60>
Additional commands
:sce <0.1>
:pi? <CR>
:pj? <CR>
:pk? <CR>
WIKA operating instructions, model CPC2000
Function
System language
Define the operating mode
Step size
Define the operating mode
Zero point adjustment
before starting the control
mode
Zero point adjustment
interval
Set communication
interface echo
Reading the current actual
pressure with unit
Reading the current actual
pressure
Reading the current unit
as text
Response
d = German
e = English
a --> Starting the automatic mode
c --> Starting the control mode
m --> Starting the measuring mode
1 ... 100 --> step size from operating the
up and down buttons in %
a --> Starting the automatic mode
c --> Starting the control mode
m --> Starting the measuring mode
(only available in CTRL and MEAS
modes)
z --> ZERO
(only available in CTRL mode)
l --> LEAK test
l --> back to CTRL mode (like [LEAK]
button)
v --> VENT (vent the entire system)
(only available in AUTOMATIC mode)
0 --> Off
1 --> On
1 ... 60 --> time in minutes
0 --> Echo Off
1 --> Echo On
On: default. Sent command is returned,
following the response, the command
will be concluded with "OK".
Off: Only the pure response will be
transmitted.
Example:
-0.05;mbar
Example:
-0.05
Example:
Mbar
EN
33