10.2 Electromagnetic Emission
Guidance and Manufacturer´s Declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product 
should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker 
emissions IEC 61000-3-3
10.3 Electromagnetic Immunity
Guidance and Manufacturer´s Declaration
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product 
should assure that it is used in such an environment.
Test / Compliance
Immunity test
Electrostatic 
± 8 kV contact
discharge (ESD)
± 2 kV, ± 4 kV, ± 8 kV, 
IEC 61000-4-2
± 15 kV air
± 2 kV for power 
supply lines; 100 kHz 
Electrostatic 
repetition frequency
transient / burst
± 1 kV for input / 
IEC 61000-4-4
output lines; 100 kHz 
repetition frequency
Surge 
± 1 kV line to line
IEC 61000-4-5
± 2 kV line             t o earth
< 0% U
Voltage dips, 
short 
at 45° steps
interruptions 
0% U
and voltage 
variations on 
70% U
power supply 
cycles
input lines
< 5% U
IEC 61000-4-11
cycles
Power 
frequency 
(50/60 Hz) 
30 A/m
magnetic             f ield
IEC 61000-4-8
12
Compliance
Group I
Class B
Class A
Complies
Electromagnetic environment – guidance
level
Floors should be wood, concrete or ceramic tile. If floors are covered               w ith 
synthetic material, the relative humidity should be at least             3 0 %.
Mains power quality should be that of a typical commercial or hospital 
environment.
Mains power quality should be that of a typical commercial or hospital 
environment.
 for 0.5 cycle 
T
Mains power quality should be that of a typical commercial or hospital              
 for 1 cycles
environment. If the user of this product requires continued operation               d uring 
T
power mains interruptions, it is recommended that the product             i s powered from 
 for 25 / 30 
T
an un-interruptible power supply or a battery.
U
 is the a. c. mains voltage prior             t o application of the test level.
T
 for 250 / 300 
T
Power frequency magnetic fields should be at levels characteristic             o f a typical 
location in a typical commercial or hospital environment.
Electromagnetic environment - guidance
This product uses RF energy only for its internal 
function. Therefore, its RF emissions are very low 
and are not likely to cause any interference in 
nearby electronic equipment.
This product is suitable for use in all 
establishments, including domestic 
establishments and those directly connected to 
the public low-voltage power supply network 
that supplies buildings used for domestic 
purposes.
1630000-F