9.2 Control Unit
Main Supply
Rated Input Current
Supply Fuse
Noise Level
Classification
Cycle Time
Size
Weight
Air Flow
Operating Pressure
Power
Control unit fuse
Ingress protection
* Protected from touch by fingers and objects greater than 12 millimeters. Protected against vertically falling drops of water or 
condensation.
9.3 Materials
Foam
Cover
Air Cells
Glide Membrane
Control unit and mattress components do not contain natural rubber latex.
10 Electromagnetic Compatibility
10.1 General EMC Information
Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual.
This product has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2 for Class B equipment.
Portable and mobile RF communications equipment can affect the operation of this product. 
Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standard. 
To determine if the emission from this product is causing the interference, run and stop running this product. If the interference 
with the other device operation stops, then this product is causing the interference. In such rare cases, interference may be reduced 
or corrected by the following:
 
Reposition, relocate, or increase the separation between the devices.
l
 
11
220 – 240 V~, 50/60 Hz
1 A
1 A
≤ 24 dB
Class II Type BF
10 min, A/B +/- 1 min
275 mm × 155 mm × 105 mm
1.75 kg
4 L/min
60 mmHg (8 kPa)
23 VA
T1 AL 250 V
IP21 *
Combustion-modified polyether polyurethane foam
Polyurethane-coated nylon fabric
Polyurethane-coated nylon 
Polyurethane film
10 Electromagnetic Compatibility
1630000-F

Werbung