Fault
The eluted antibody has
low purity.
Cause
Cross-contamination
from a previous purifica-
tion
The washing step after
sample loading was in-
sufficient to remove con-
taminants.
8
Disposal
8 . 1
Decontaminating the Product
Procedure
If the product has come into contact with hazardous substances, decon-
t
taminate the product.
8 .2
Disposing of the Product
The product must be disposed of properly by disposal facilities.
Requirements
If applicable: The product must be decontaminated.
Procedure
Dispose of the product and packaging in accordance with local govern-
t
ment regulations.
Solution
Perform a regeneration step after each
purification cycle.
When purifying antibodies from multiple
sources, use a separate product for each
antibody type and | or batch.
Increase the buffer volume and | or re-
duce the flow rate during washing.