Operation
16
Instructions for Use Sartobind
Requirements
The pump tubing is loaded into the peristaltic pump and filled with equili-
bration buffer .
Procedure
Run the peristaltic pump at a low flow rate .
t
Hold the product in an upright position with the arrow indicating the flow
t
direction (2) pointing down.
1
Remove the protective cap (1) from the product inlet.
t
2
If operating without inline prefiltration:
t
Connect the Luer-Lock connector (3) on the pump tubing to the
t
product inlet (4) drop-to-drop .
3
4
If operating with inline prefiltration:
t
Connect the Luer-Lock connector (5) on the pump tubing to the mi-
t
crofilter inlet .
Fill the microfilter (6) with buffer via the pump.
t
5
Connect the microfilter outlet to the product inlet (7) drop-to-drop.
t
6
7
Remove the protective cap from the product outlet.
t
If air was introduced into the product:
t
Invert the product so that the arrow indicating the flow direction (8)
t
points up .
Rinse the product with equilibration buffer until no air bubbles
t
emerge from the product outlet.
8
Return the product to an upright position with the arrow indicating
t
the flow direction pointing down.
Flush the product with 10 mL equilibration buffer at up to 25 mL/min.
t
The product is equilibrated and ready for sample loading.
y
Rapid A Lab
®