Operation
18
Instructions for Use Sartobind
5 .4
Using a LC System
5 .4 . 1
Preparing the LC System
Materials required:
Procedure
Prepare the LC system for purification (see instructions for the LC sys-
t
tem).
Connect the UNF 10|32 adapters to the LC system. Make sure the con-
t
nections are fluid-tight.
5 .4 .2
Installing and Equilibrating the Product
Materials required:
Requirements
The UNF 10|32 adapters are connected to the LC system.
Procedure
Flush the LC system with equilibration buffer.
t
Make sure the flow path is free from air.
t
Hold the product in an upright position with the arrow indicating the flow
t
direction (2) pointing down.
1
Remove the protective cap (1) from the product inlet.
t
2
If operating without inline prefiltration:
t
Run the LC system at a low flow rate (~ 1 mL/min).
t
Disconnect the UNF 10|32 adapters from each other.
t
Connect the upstream UNF 10|32 adapter (3) to the product inlet
t
drop-to-drop.
3
Rapid A Lab
®
Buffers, see chapter "5.4.3 Performing a Purifica-
tion Cycle", page 20
— Optional: 0.2 µm microfilter for inline prefiltra-
tion (see chapter "9.7 Approved Microfilters",
page 25)
— > 10 mL equilibration buffer, e.g., 1 x PBS pH 7.4