Please pay attention to the following important points for the maintenance of the material:
Clean, disinfect and sterilize the device and components after each use.
Do not use dissolving agents for cleaning.
Autoclave material in transparent packaging.
Do not fill the sterilization bag more than 80%.
Autoclave material at maximum 134° C.
Sterilized material should be stored and tagged with sterilization date,

6.1 Control unit and foot control

The control unit and the foot control do not come in contact with patients. Clean the surface of the unit
only, do not use harsh cleaners or solvents for cleaning. Use 80% ethyl alcohol or microbiologically
effective disinfectants. The control unit face is sealed and washable.

6.2 Electronicmotor 31ESS

To avoid breakage, do not bend motor cable!
Do not clean the electronicmotor with compressed air!
Clean, disinfect and sterilize the electronicmotor after each use!
Autoclaving the electronicmotor without packing it into a transparent bag may
seriously damage it!
After autoclaving, allow electronicmotor to dry for one hour at room
temperature.
1)
The electronicmotor, cable and plug should be cleaned after each use to prevent the build up of
deposits and debris that can destroy it, if not removed. Wipe it with a clean cloth dampened with
disinfectant solution.
2)
Spray the motor interior with Nou-Clean - spray.( See spray can instructions).
3)
Pack Motor in transparent bag (see DIN 58953).
4)
After packing motor and cable, autoclave it at maximum 134° C .
After autoclaving, allow electronicmotor to dry for one hour at room temperature.

6.3 Tubing set Nr. 1706

Disposable tubing set Nr. 1706can not be sterilized!

6.4 Y-Connector

After use, flush the saline residue out of the Y-Connector with distilled water for
approximately 20 seconds with distilled water!
1)
Flush the Y-Connector with connected contra angle with distilled water for approximately 20
seconds.
2)
Remove the tubings from the Y-Connector.
3)
Pack the Y-Connecotor in a transparent bag and sterilize the pack in an autoclave at max. 134° C.
No. 31820 22/07

6 Disinfection, Cleaning and Sterilization

MD 10
9