TROUBLESHOOTING

9.1
Diagnostic System and Symbols on Keyboard
PIEZOSURGERY® flex is equipped with a diagnostic system able to detect operating anomalies.
Icons are displayed on the touchpanel according to the detected operating anomalies.
By using the following chart, the user is guided toward the identification and possible solution
of the malfunction detected.
Symbol on the Keyboard
34
flex
WARNING: Portable RF communication
equipment (including peripheral devices as
antenna cables and external antennas) must
not be used closer than 30 cm (12 inches) to
any part of the device PIEZOSURGERY® flex,
including the cables specified by the
manufacturer. otherwise, there may be a
performance degradation of these devices.
Possible Cause
Handpiece electric contacts/
lead wet.
Piezosurgery® Medical
handpiece not connected to
the device.
Faulty Handpiece.
Malfunction of the tuning
circuit.
Insert not properly tightened
on the handpiece.
Insert broken, worn or
deformed.
Handpiece electric contacts/
lead wet.
Solution
Thoroughly dry the contacts
with compressed air.
Connect the handpiece.
Replace the handpiece.
Contact an Authorized
MECTRoN Service Centre.
Unscrew the insert and screw
it back correctly using the
torque wrench (see
Chapter
5.5 on page 22
).
Replace the insert.
Thoroughly dry the contacts
with compressed air.