Problem
During
operation,
irrigation flows out from the
insert.
The device works properly
but the pump strains/ is
stressed.
The pump turns correctly, but
when it stops, irrigation flows
out from the handpiece.
Poor performance
Tabella 4 – Troubleshooting Quick Guide
Possible Cause
The insert is clogged.
no
The handpiece is clogged.
The liquid irrigation bag is
empty
The air inlet on the spike has
not been opened.
Irrigation kit not positioned
correctly inside the pump.
Excessive pressure of the
impeller on the peristaltic
pump tube.
The lid of the peristaltic
pump is not closed correctly.
The insert is not correctly
tightened on the handpiece
Broken, worn-out or
deformed insert.
TRoUBLESHooTING
Solution
Unscrew the insert from
the handpiece and unclog
the insert water passage
by blowing compressed air
through it. If the problem
persists, replace the insert
with a new one.
Contact an Authorised
Mectron Service Centre.
Replace the empty bag with
a new one full.
open the air inlet on the
spike/roller.
Check that the Irrigation kit
is properly connected.
Check that the tube inside
the peristaltic pump is
properly positioned
(Refer to
Chapter 4.2 on
page 12
)
Check that the lid of the
peristaltic pump is perfectly
closed
(Refer to
Chapter 4.2 on
page 12
)
Unscrew the insert and screw
it back correctly by using the
Mectron torque wrench.
(Refer to
Chapter 5.5 on
page 22
)
Replace the insert with a new
one.
37