TECHNICAL SPECIFICATIONS
Device compliant with Regulation (EU)
2017/745
Classification as per IEC/EN 60601-1
Essential performance
Device for intermittent operation
Power supply voltage
Max. power consumption
Fuses
Operating frequency
Power settings
Modes
Peristaltic pump capacity
APC circuit protections
Operating conditions
Transport and storage conditions
Altitude
Weight and size
Table 2 – Technical Specifications
a) W = width ; L = length ; H = height
26
flex
Classe IIa
I
Applied part type B (insert)
IP 20 (device)
IP X8 (foot pedal model FS-06)
According to the standard IEC 80601-2-60
the device has no essential performance.
60sec. ON - 30sec. OFF with irrigation
100-240 V~ 50/60 Hz
120 VA
Type 5 x 20 mm, T 2AL, 250V
Automatic scan
From 24 KHz to 36 KHz
Adjustable on the touch screen:
7 power levels, from 1 to 7
Adjustable on the touch screen:
from 1 to 3
Adjustable on the touch screen:
6 flow levels, from 1 to 6 (from 8 ml/min to
approximately 65ml/min)
No handpiece detected
Cord interruption
Insert not tightened correctly or broken
from 10 °C to 35 °C
Relative humidity from 30% to 75%
Air pressure P: 800hPa/1060hPa
from -10 °C to 60 °C
Relative humidity from 10% to 90%
Air pressure P: 500hPa/1060hPa
Less than or equal to 2000 meters
3,2Kg
300 x 250 x 95 mm (L x l x H)
a)