WARNING: Verify that the irrigation
clamp is in the 'Open' position before
operation, and that it is closed before
disconnecting the irrigation tubing from the
bag at the end of the procedure.
CAUTION: Never force the connector
into the console port as this may damage the
connector and/or console. If the connector
and the port do not join with reasonable
ease, they probably don't match. Make sure
that the marked dot on the handpiece-cord
connector is facing upwards.
CAUTION: Check that the
PIEZoSURGERY® Medical handpiece is
correctly connected before using the system.
WARNING: Before every treatment,
make sure that the insert appropriate
for the treatment is securely attached
to the handpiece. Use exclusively the
PIEZoSURGERY® Medical torque wrench to
securing the insert to the handpiece. Do not
use any other tool such as pliers, pincers, etc.
WARNING: Before every treatment,
make sure that the selected insert has been
correctly screwed onto the handpiece. This
happens when the PIEZoSURGERY® Medical
torque wrench used to tighten the insert
emits a mechanical "CLICK" sound.
CAUTION: Hold only the connector when
connecting/disconnecting the foot pedal.
Never actually pull on the cord itself.
CAUTION: Do not attempt to screw
or twist the connector during insertion or
removal: the connector could get damaged
by twisting.
CAUTION: The foot pedal is specifically
designed to be used only in connection with
the PIEZOSURGERY® flex device. Only use
an original foot pedal otherwise damages or
malfunctions can happen.
20
flex
CAUTION: Do not activate the handpiece
while the insert is in contact with the part
to be treated. Doing so will not allow the
electronic control circuit of the console to
recognise the best point of resonance of
the insert required for efficient, optimum
performance.
WARNING: The patient must not come
into contact with the device body or the foot
pedal.
WARNING: Single use - Before the
surgery. Before using any PIEZoSURGERY®
Medical insert inspect the sterile package
and the product for any damage. Insert
loses sterility in case of broken or damaged
packaging. In case of damaged packaging,
the insert MUST be disposed of. DO
NOT reprocess or reuse the insert whose
packaging is damaged". Before starting the
surgery correctly tighten the insert onto the
handpiece by means of the torque wrench.
WARNING: Do not change the insert
while the handpiece is operating to prevent
possible user injury.
WARNING: During the tightening and
removing operations, the user must pay
particular attention to avoiding injury from
inserts with sharp points and cutting edges.