Technical data
!
32
9
Technical data
9.1
Power supply
Regeneration unit: 230 V; 50 Hz; 2100 W
Compartment: no power supply required!
9.2
Operating conditions
The typical cooling period for a completely regenerated cooling compartment at 30°C ambient
temperature and 16°C medium temperature is 48 hours. For example 16 l sample medium will be
cooled to 4°C within 8 hours and held at that temperature under these conditions.
Ambient temperature: 0 to 40°C
Storage temperature: -20 to 60°C
Note!
• Do not expose the compartment to direct sunlight!
• Do not set up the regeneration unit in the open air!
9.3
Mechanical construction
Dimensions:
480 mm
521 mm
Fig. 7:
Dimensions of Liquiport 2000 with active cooling compartment
Weight of compartment: 37 kg
Weight of regeneration unit: 45 kg
Dimensions of regeneration unit (L x B x H): 870 x 700 x 994 mm
Liquiport 2000 Active cooling
Endress+Hauser