Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting - EKOM DK50 2V MOBILE MINI Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

TROUBLESHOOTING

17.9. Check of fan
It is necessary to keep the entire device, especially the compressor and cabinet fans, clean. Remove
dust from the fans using vacuum or compressed air.
17.10. Cleaning and disinfection of the exterior surfaces of the product
Use neutral agents for cleaning of the external surfaces of the product.
The use of aggressive cleaning agents and comprising alcohol and chlorides
may lead to the damage of the surface and the discoloration of the product.
TROUBLESHOOTING
Risk of electric shock.
Before any of the following operations on the device, disconnect the device from
the mains (pull out the mains plug).
Risk of injury during work with pneumatic components under pressure.
Before any of the following operations on the device, it is necessary to decrease
pressure in the air tank and in the pneumatic system to zero.
Any operations concerning the troubleshooting may be performed only by
a qualified technician.
Malfunction
Compressor does
not switch on
Compressor often
switches on
Low pressure in
the air tank
(compressor
running constantly)
10/2019
Possible cause
No voltage in the pressure switch
Motor winding failure, damaged
thermal protection
Capacitor failure
Seized piston or other rotating part
Pressure switch does not switch
Air leakage in pneumatic system
Non-return valve (SV) leakage
Large volume of condensed liquid in
pressure tank
Low compressor capacity
High air consumption of supplied
equipment
Leakage in pneumatic system
Solution
Check voltage in socket
Check circuit breaker switch –
switch to position switched-on „I"
Loose conductor from terminal -
repair
Check electrical cord – replace
defective cord
Replace motor or windings
Replace capacitor
Replace damaged components
Check function of pressure switch
Check pneumatic system – seal
loose connections
Clean non-return valve, replace
seals, replace non-return valve
Drain condensed liquid
Check time of filling the air tank
Decrease air consumption
Use compressor with higher
capacity
Check pneumatic system – seal
loose connections
27
NP-DK50 2V MOBILE mini-1_11-2019

Werbung

Inhaltsverzeichnis
loading