Fig. 3: Releasing the pump
Once the panel is installed, the grounding wire from the cabinet panel must be connected and
the cabinet must then be closed and secured with screws.
Connect devices to the compressed air outlet on the cabinet.
Insert the mains plug into a rated outlet circuit.
11. PNEUMATIC CONNECTION
11.1. Connecting to the compressed air outlet
There is a G1/4" outlet with internal thread for the connection of pressure hose in the rear upper
part of the cabinet.
Risk of damage to pneumatic components.
Air hoses must not be broken.
12. ELECTRICAL CONNECTION
The product is delivered with a cord equipped with a plug with earthing pin.
Keep the socket easily accessible to ensure that the device can be safely disconnected from the
mains.
The corresponding circuit must be protected in the power distribution max. 16 A.
Risk of electric shock.
It is necessary to follow all local electrotechnical regulations. The mains voltage
and frequency must comply with the data stated on the device label.
Risk of fire and electric shock.
Electrical cable must not be in contact with hot compressor components.
Risk of fire and electric shock.
Electrical cord must not be broken.
NP-DK50 2V MOBILE mini-1_11-2019
18