Herunterladen Inhalt Inhalt Diese Seite drucken

Rs-232 Interface (Option); Connections - Philips PM 5330 Handbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

GB
4 - 16
CHARACTERISTICS
PM 5330
4.11
RS-232 INTERFACE (Option)
command set
compatible with IEEE-488.2
operating modes
communication mode
baud rates
110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200
data bits
7 or 8
stop bits
1 (2 for 110 baud only)
parity
odd, even or no (with 8 data bits)
X-on/X-off handshake
on or off
isolation
in-/outputs galvanically separated with opto-couplers
hardware connection
3 wires, no hardware handshake
7 wires, with hardware handshake
hardware handshake
DSR/DTR and CTS/RTS
connector
9-pin D-connector (male)
4.12
CONNECTIONS
front panel
rear side
RF OUT Zo 50 S2
MOD IN
COUNTER IN
IEEE-488/RS-232
MOD OUT
SWEEP OUT
REF IN
BNC socket, RF signal output
BNC socket,
input for ext. modulation signals
BNC socket,
input for frequency counter
standard interface connector for
IEEE-488 bus or 9-pin D-connector
for RS-232
BNC socket, Zo = 10 kS2, internal
modulation signal 1 V(pp) sine wave
for AM and FM INT, stereo
BNC socket, sweep voltage
proportional to frequency
0 V =centre frequency
- 5 V =lower frequency
+ 5 V =upper frequency
BNC socket,
for external synchronization,
see Section 4.3.2
10 MHz OUTPUT
BNC socket, int. reference signal,
see Section 4.3.2

Werbung

Inhaltsverzeichnis
loading