INSTALLATION- AND SAFETY INSTRUCTIONS
PM 5330
1.1.3
Mains Voltage Setting and Fuses
Before inserting the mains plug into the mains socket, make sure that the instrument is set to
the local mains voltage.
WARNING: If the mains plug has to be adapted to the local situation, such adaption
should be done by a qualified person only.
On delivery from the factory the instrument is set to one of the following mains voltages.
Instrument
version
Code no.
Mains voltage
setting
delivered
mains cable
PM 5330/..1
9452 053 30. .1
220 V
Europe, Schuko
PM 5330/..3
9452 053 30. .3
120 V
North America
PM 5330/..4
9452 053 30. .4
240 V
England (U.K.)
PM 5330/. .5
9452 053 30..5
220 V
Switzerland
PM 5330/..8
9452 053 30. .8
240 V
Australia
The set mains voltage and the corresponding fuse are indicated on the rear panel .
Make sure that only fuses of the required current rating, and of the specified type, are used for
renewal . The use of repaired fuses, and / or the short -circuiting of the fuse holder, are pro-
hibited.
The fuse shall be renewed only by a qualified person who is aware of the hazard involved.
WARNING: The instrument shall be disconnected from all voltage sources when a fuse
is to be replaced or when the instrument is to be adapted to a different
mains voltage.
The instrument can be set to the following mains voltages: 100 V, 120 V, 220 V and 240 V a.c.
These nominal voltages can be selected by means of the mains-voltage selector, located at
the rear panel , adjacent to the mains socket. The fuse is located in a holder at the same place.
For mains voltage selection or replacement of the fuse remove the mains cable and pry open
the fuse compartment with a small screwdriver (see drawing) .