13.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1 x short
2 x short
1 x long
Switching the Axon-Bus prosthetic system on/off
Charging recept
acle
/
/
13.2.2 Warnings/error signals
Error during use
Charging recept
acle
-
-
Error while charging the product
Battery charger
Event
Battery charger connected
Charging begins
Charging ends
Beeper
2 x short
1 x long
Beeper
1x long
-
3 x short
Pulsating
Explanation/correction
Axon-Bus prosthetic system is not charging
•
Disconnect and then reconnect the charging plug.
•
Clean charger contacts.
Charger is overheated or overcooled
•
Disconnect charging plug and allow charger to cool
down/warm up.
Event
Switch on the Axon-Bus prosthetic system
Switch off the Axon-Bus prosthetic system
Explanation/correction
Error in the Axon-Bus prosthetic system
•
Switch the Axon-Bus prosthetic system off/on
•
Open/close the Axon-Bus prosthetic system
•
Consult prosthetist if error persists
Axon-Bus prosthetic system outside operat
ing temperature range
•
Allow the Axon-Bus prosthetic system to
warm up or cool down (Technical data)
Axon-Bus prosthetic system switches itself
off
•
Charge battery
Opening the Axon-Bus gripping component
in an emergency
•
System is switched off, system can be used
without restriction after being switched on
again