Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850/GSM 900: 0.50 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.35 m
• DECT cordless phones incl. base station: 0.18 m
• WiFi (routers, access points,...): 0.11 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.11 m
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the product.
CAUTION
Remaining in areas outside the allowable temperature range
Injury due to faulty control or malfunction of the product.
Avoid remaining in areas outside the allowable temperature range (see Page 25).
CAUTION
Penetration of product with dirt and humidity
Injury due to unexpected product behaviour or malfunction.
Ensure that neither solid particles nor liquids can penetrate into the product.
CAUTION
Overloading due to unusual activities
Injury due to unexpected product behaviour as a result of a malfunction.
The product was developed for everyday use and must not be used for unusual activities.
These unusual activities include, for example, sports with excessive strain and/or shocks to
the wrist joint (pushups, downhill mountain biking, ...) or extreme sports (free climbing,
paragliding, etc.).
Careful handling of the product and its components not only increases their service life but,
above all, ensures the patient's personal safety!
If the product and its components have been subjected to extreme loads (e.g. due to a fall,
etc.), then the product must be inspected for damage immediately. If necessary, forward the
product to an authorised Ottobock Service Centre.
CAUTION
Changing Axon-Bus gripping components when switched on
Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.
Power down the Axon-Bus prosthetic system by pressing the button in the charging recept­
acle before changing any Axon-Bus components (e.g. Axon-Bus gripping component).