9 Cleaning
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).
1) Clean the product with a damp cloth and mild soap (e.g. Ottobock 453H10=1 Derma Clean)
when needed.
Ensure that no liquid penetrates into the system component(s).
2) Dry the product with a lint-free cloth and allow it to air dry fully.
10 Maintenance
10.1 Replacing hook tip
Required Tools
710418=13 Crowfoot spanner
9S305 hook tip (set)
Removing the hook tip
1) Slide the cap nut cover over the hook tip and remove it.
2) Loosen the cap nut with the spanner (see fig. 1).
3) Unscrew the cap nut (see fig. 2) and pull off the hook tip with the cap nut.
Mounting the hook tip
1) Place the hook tip on the connector of the gearbox (see fig. 4).
INFORMATION: Mount the hook tip so that the groove of the hook tip lies parallel over
the pin of the connector (see fig. 3).
2) Screw on the hook tip.
3) Tighten the hook tip with the spanner (see fig. 5).
INFORMATION: The hook tip must be tightened using a torque of 17 Nm.
4) Pull the cap nut cover over the hook tip and position it over the cap nut (see fig. 6).
10.2 Initialising the Axon-Bus gripping component
The Axon-Bus gripping component is delivered in an initialised state.
CAUTION
Risk of pinching between the hook tips
Injury due to pinching of body parts.
Ensure that no body parts are between the hook tips when using or rotating the product.
24
>
If the two hook tips no longer close completely, the following steps
must be carried out for initialisation:
1) Remove the AxonHook from the socket while opening the hook tips
(see Page 22).
2) Switch off the Axon-Bus prosthetic system (see Page 22).
3) Reattach the AxonHook to the socket.
4) Switch the Axon-Bus prosthetic system back on.
5) Open the hook tips completely.
→ The hook tip is opened at low speed.
6) Close the hook tips completely.
→ The hook tip is closed at low speed.