Herunterladen Inhalt Inhalt Diese Seite drucken

Warnings/Error Signals - Ottobock Kenevo 3C60 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Kenevo 3C60:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Beep signal
1 x short

13.2.2 Warnings/error signals

Error during use
Beep signal
10 x long
30 x long
62
Vibration signal
1x before beep signal Battery charger disconnected after start of char­
Vibration signal
1 x long at interval of
approx. 5 seconds
3 x long
5 x long
10 x long
1 x long, 1 x short
repeated every 3
seconds
Event
ging mode
Event
Overheated hydraulic
unit
Battery charge level
under 35%
Battery charge level
under 20%
Battery charge level
under 15%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
For example, sensor
not operational or
valve drive failure
Possibly no switching
into safety mode.
Required action
Reduce activity.
Charge battery soon.
Charge battery immedi­
ately; the product will be
switched off after the next
warning signal.
Charge the battery.
Walking possible with
restrictions. Please note
the possible change in
flexion/extension resist­
ance.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger. The battery charger
must remain connected
for at least 5 seconds
before it is disconnected.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Kenevo 3c60-st

Inhaltsverzeichnis