Beep
Vibration
signal
signal
3 x short
Battery charge level
Charger
Battery is charging.
The on time of the LED indicates the current battery charge level.
The on time of the LED gets longer as the charge level increases. It only flashes
briefly at the start of the charging process and stays on continuously at the end of
the charging process.
Battery is fully charged, or the temperature fell above/below the permissible range
during charging. Check current charge level (see Page 48).

13.3 Directives and manufacturer's declaration

13.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see
Page 42).
The following tables outline the test levels of the tests to be carried out. The higher value applies if
there is a difference between the operating environments listed in the tables below.
Table 1 – Electromagnetic emissions for all devices and systems
Interference measurements
HF emissions according to
CISPR 11
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
64
Event
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery
charger. If the beep signal sounds again, you should visit your prosthet­
ist soon. If necessary, the prosthetist will forward the product to an
authorised Ottobock Service Centre.
The product can be used without restrictions. However, vibration signals
may not be generated.
Compliance
Group 1
Class B
N/A
N/A
Electromagnetic environ­
ment directive
The product uses HF energy
exclusively for its internal func­
tioning. Its HF emissions are
therefore very low, and inter­
ference with neighbouring
electronic devices is unlikely.
The product is intended for
use in other than residential
facilities and in facilities con­
nected directly to a public low-
voltage network that also sup­
plies residential buildings.