Herunterladen Inhalt Inhalt Diese Seite drucken

Deep Sleep Mode - Ottobock Genium Gebrauchsanweisung

Bionic prosthetic system
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Remote Control
Remote Control
INFORMATION
The prosthetist's setting is usually found in the middle of the scale. After making changes, the setting established
by the prosthetist can be restored by moving the slider to the middle.
INFORMATION
The knee joint settings should be optimised by the prosthetist using X-Soft. The remote control is not intended for
use by the prosthetist in order to set up the knee joint. The day-to-day behaviour of the prosthesis can be changed
to a certain degree when the patient uses the remote control (e.g. as the user is becoming accustomed to the pros-
thesis). At the next appointment, the prosthetist can trace the changes in X-Soft.
The tables below show the adjustment parameters for the individual MyModes along with a brief description, the
adjustment range of the remote control around the value selected by the prosthetist and the absolute value range.
Basic mode:
X-Soft
Remote control Absolute range
Stance phase
Resistance
flexion resistance
Swing flexion angle Angle
Pitch
Pitch
Volume
MyModes except knee braking mode:
X-Soft
Remote control Absolute range
Increase
Increase
Pitch
Pitch
Volume
Volume
Knee braking mode:
X-Soft
Remote control Absolute range
Basis
Basis
Pitch
Pitch
Volume
Volume
7.7  Deep sleep mode
INFORMATION
Advise your patient of the information in this section.
The remote control can be used to place the knee joint in a deep sleep mode, in which power consumption is mini-
mised. The knee joint offers no functionality in this mode.
It can be awakened from deep sleep mode with the remote control or by connecting the charger.
56 | Ottobock
56 | Ottobock
Remote 
adjustment range
120-180
±10
55°-70°
±3°
1000Hz-4000Hz
1000Hz-4000Hz
0-4
0-4
Remote 
adjustment range
0-100
±10
1000Hz-4000Hz
1000Hz-4000Hz
0-4
0-4
Remote 
adjustment range
0-180
±20
1000Hz-4000Hz
1000Hz-4000Hz
0-4
0-4
control 
Description
Resistance against flexion motion
e.g. when walking down stairs or
when sitting down on a chair.
Maximum flexion angle in the swing
phase.
Frequency (pitch) of the beeper for
confirmation signals.
Volume of the beeper for confirma-
tion signals.
control 
Description
Value for speed at which flexion re-
sistance increases as the knee ang-
le increases.
Frequency (pitch) of the beeper for
confirmation signals.
Volume of the beeper for confirmati-
on signals.
control 
Description
Basic resistance for the knee joint
without load
Frequency (pitch) of the beeper for
confirmation signals.
Volume of the beeper for confirmati-
on signals.
Genium - Bionic Prosthetic System
Genium - Bionic Prosthetic System

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis