Alignment and Adjustment of the Prosthesis
Alignment and Adjustment of the Prosthesis
INFORMATION
When the knee joint is fully charged (all 5 LEDs at the side light up (Fig. 3)), the maximum operating time of the
knee joint before the next charging process is up to 5 days, depending on patient activity. However, we recommend
charging the knee joint every night.
2.2  Connecting the AXON tube adapter
1. Determine the required length of 2R20/2R21 AXON Tube Adapter using the X-Soft adjustment software.
2. Shorten the AXON tube adapter to the determined value with the 719R3 Tube Adapter Cutter (Fig. 1). Smooth
the cutting surface with a 718S2 Deburring Knife and sandpaper (Fig. 2). If necessary, smooth by machine the
material raised at the outer edge when shortening the tube adapter, and always deburr the interior to prevent
damage to the tube adapter cable.
3. Finally, the plug cover must be removed.
 CAUTION
Damage to the cable. If the cable is damaged when cutting the tube adapter to length, malfunctions of the knee
joint may result. This can cause the patient to fall.
Ensure the cable is not damaged.
INFORMATION
Do not cut the tube adapter below the 365 mm value on the tube adapter scale.
INFORMATION
If the tube adapter is disconnected when the knee joint is activated and the charger is not connected, safety mode
is activated and the knee joint sends a warning signal. If the tube adapter is connected again and then the knee
joint is reset by attaching and removing the charger, it returns to normal mode.
INFORMATION
If the knee head is fully flexed while the joint is standing upright, the joint switches to sitting or energy-saving mode
after 2 seconds. Then the tube adapter can be disconnected without triggering an error message. Ensure that there
is no active Bluetooth connection.
This function can be used for storage, shipping or to replace tube adapters.
4. Attach the foot to the tube adapter. To do so, tighten the screws on the tube adapter slightly with an Allen key (size
4 mm) (approx. 4 Nm). Fit the tube adapter to the foot so that the printed scale faces the front (Fig. 3).
5. Attach the cable of the tube adapter to the cable of the 3B1/3B1=ST Knee Joint (Fig. 4).
48 | Ottobock