Herunterladen Inhalt Inhalt Diese Seite drucken

Adjusting The Torsion Moment On The 2R21 Axon Tube Adapter; Bench Alignment In Alignment Apparatus; Static Alignment Optimisation; Dynamic Alignment Optimisation - Ottobock Genium 3B1-2 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Genium 3B1-2:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Preparation for Use

7.1.3 Adjusting the torsion moment on the 2R21 AXON tube adapter

CAUTION
Incorrect setting of the torsion moment in the torsion unit
Falling due to unexpected behaviour of the product.
► The marking on the Allen head screw may not be turned as far as the red area or beyond the red area.
The torsion moment can be adjusted with the Allen head screw in the centre of the adapter.
Increasing the torsion moment:
► Turn the mark in the centre of the torsion unit clockwise.
Decreasing the torsion moment:
► Turn the mark in the centre of the torsion unit counterclockwise.

7.1.4 Bench alignment in alignment apparatus

With correct bench alignment, e.g. in the PROS.A. assembly alignment apparatus (743A200), the benefits of the
product are realised to best advantage. If the L.A.S.A.R. assembly alignment apparatus (743L200) is available, it
can be used as well.
Alignment can also be carried out using LaserLine/plumb line.
A calibrated alignment recommendation for prosthesis bench alignment is provided in the adjustment software for
the individual conditions of the prosthesis and patient. Therefore, consult the adjustment software for the alignment
data.
The following points must be observed during alignment:
Static alignment in the alignment apparatus must always be carried out without shoes, otherwise a correct
adjustment will not be possible.
Static alignment with the LaserLine/plumb line must be carried out with shoes, otherwise a correct adjust­
ment will not be possible.
Ensure the knee joint is in full extension during bench alignment. To do so, briefly push the socket once into the
fully extended position.

7.1.5 Static alignment optimisation

The adjustment software specifies concrete reference values based on measurement data to help optimise the
alignment.
A precondition is that the alignment recommendations were followed during bench alignment of the prosthesis.
The goal of optimised alignment is that the user must compensate as little as possible with their residual limb.
Optimised alignment of the prosthesis components can reduce the effort required from the patient.
INFORMATION
During static alignment optimisation, the knee joint is automatically locked in the flexion direction. This should
enable the patient to stand in a stable stance that is unaffected by the alignment. Patients can only walk with the
prosthetic leg fully extended in this situation!

7.1.6 Dynamic alignment optimisation

After adjusting the product with the adjustment software, perform dynamic optimisation during trial walking. Often,
the following aspects have to be observed and adapted, if necessary:
Socket flexion position by verifying step length symmetry (sagittal plane)
Adduction position of the socket and M-L positioning of the socket adapter (frontal plane)
Rotation position of the knee joint axis and outward rotation of the prosthetic foot (transversal plane)

7.1.7 Torque values of the screw connections

CAUTION
Incorrectly secured screws
Falling due to breakage of load-bearing components caused by screw connections coming loose.
► After completing all settings, the set screws in the tube adapter must be secured before they are tightened to
the specified torque.
► The screws in the clamp bracket must not be secured but only tightened to the specified torque.
Using the 710D4 torque wrench, tighten the corresponding screws alternately in several cycles until the specified
torque is reached.
68 | Ottobock
Genium 3B1-2/3B1-2=ST

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Genium 3b1-2 st

Inhaltsverzeichnis