Herunterladen Inhalt Inhalt Diese Seite drucken

Inspection, Maintenance And Checks; Steam Sterilization - B. Braun Aesculap Gebrauchsanweisung/Technische Beschreibung

Activo-rahmensystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 30

Inspection, maintenance and checks

Damage (metal seizure/fric-
tion corrosion) to the product
caused by insufficient lubrica-
CAUTION
tion!
Allow the product to cool down to room tempera-
ture.
After each complete cleaning, disinfecting and dry-
ing cycle, check that the product is dry, clean,
operational, and free of damage (e.g. broken insu-
lation or corroded, loose, bent, broken, cracked,
worn, or fractured components).
Dry the product if it is wet or damp.
Repeat cleaning and disinfection of products that
still show impurities or contamination.
Check that the product functions correctly.
Immediately put aside damaged or inoperative
products and send them to Aesculap Technical Ser-
vice, see Technical Service.
Check for compatibility with associated products.
Prior to function checks,
lubricate moving parts (e.g.
joints, pusher components
and threaded rods) with
maintenance oil suitable
for the respective steriliza-
tion process (e.g. for steam
sterilization: Aesculap
STERILIT® I oil spray JG600
or STERILIT® I drip lubrica-
tor JG598).
Packaging
Place the product in its holder or on a suitable tray.
Ensure that all cutting edges are protected.
Pack trays appropriately for the intended steriliza-
tion process (e.g. in Aesculap sterile containers).
Ensure that the packaging provides sufficient pro-
tection against recontamination of the product
during storage.

Steam sterilization

Note
The product can be sterilized either in disassembled or
in assembled condition.
Check to ensure that the sterilizing agent will come
into contact with all external and internal surfaces
(e.g. by opening any valves and faucets).
Validated sterilization process
– Steam sterilization using fractionated vacuum
process
– Steam sterilizer according to DIN EN 285 and
validated according to DIN EN ISO 17665
– Sterilization using fractionated vacuum pro-
cess at 134 °C/holding time 5 min or as per
locally approved policies and procedures
When sterilizing several products at the same time
in a steam sterilizer, ensure that the maximum load
capacity of the steam sterilizer specified by the
manufacturer is not exceeded.
25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis