Aesculap
activO frame system
Manual pre-cleaning with ultrasound and brush
Phase
Step
I
Ultrasonic cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath
(frequency 35 kHz) for at least 15 min. Ensure that
all accessible surfaces are immersed and acoustic
shadows are avoided.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
22
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
Phase II
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
D–W
-
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.