Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested
and approved effectiveness (e.g. FDA approval or CE
mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for process-
ing must be serviced and checked at regular intervals.
Manual pre-cleaning with a brush
Phase
Step
I
Disinfec-
tant
cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfec-
tant for at least 15 min. Ensure that all accessible
surfaces are moistened.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
T
t
[°C/°F]
[min]
RT (cold)
>15
RT (cold)
1
Conc.
Water
[%]
quality
2
D–W
-
D–W
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Chemical
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
-
21