activO frame system
Note the information on appropriate cleaning
brushes and disposable syringes, see Validated
cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfec-
tant for at least 15 min. Ensure that all accessible
surfaces are moistened.
Clean the product with a suitable cleaning brush in
the solution until all discernible residues have been
removed from the surface.
If applicable, brush through non-visible surfaces
with an appropriate cleaning brush for at least
1 min.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Drain any remaining water fully.
16
Phase III
Fully immerse the product in the disinfectant solu-
tion.
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Rinse lumens at least 5 times at the beginning of
the exposure time using an appropriate disposable
syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible
surfaces).
Mobilize non-rigid components, such as set screws,
joints, etc. during final rinse.
Rinse lumens with an appropriate disposable
syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.