Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with immer-
sion disinfection
■
For BV399R, BV842R,
BV862P, BV863P, BV864P,
BV868R, BV859R,
BV869R, BV899R, FK166R,
RT092R, RT069R, and
OP803
Manual cleaning with ultra-
sound and immersion disin-
fection
■
For BV814R/BV815R and
BV843R
Mechanical alkaline cleaning
and thermal disinfection
■
For BV399R, BV859R and
FK166R
Mechanical neutral or mild
alkaline cleaning and thermal
disinfecting
■
For OP803
Specific requirements
■
Suitable cleaning brush:,
e.g. TA011944
■
20 ml disposable syringe
■
When cleaning products with movable
hinges, ensure that these are in an
open position and, if applicable, move
the joint while cleaning.
■
Drying phase: Use a lint-free cloth or
medical compressed air
■
Suitable cleaning brush:,
e.g. TA011944
■
20 ml disposable syringe
■
When cleaning products with movable
hinges, ensure that these are in an
open position and, if applicable, move
the joint while cleaning.
■
Drying phase: Use a lint-free cloth or
medical compressed air
■
Place the product in a tray that is suit-
able for cleaning (avoiding rinsing
blind spots).
■
Place products in the tray with their
hinges open.
■
Place the product in a tray that is suit-
able for cleaning (avoiding rinsing
blind spots).
■
Connect components with lumens and
channels directly to the rinsing port of
the injector carriage.
■
Keep working ends open for cleaning.
■
Place products in the tray with their
hinges open.
Reference
Chapter Manual cleaning/disin-
fection and sub-chapter:
■
Chapter Manual cleaning
with immersion disinfection
Chapter Manual cleaning/disin-
fection and sub-chapter:
■
Chapter Manual cleaning
with ultrasound and immer-
sion disinfection
Chapter Mechanical cleaning/
disinfecting and sub-chapter:
■
Chapter Mechanical alka-
line cleaning and thermal
disinfecting
Chapter Mechanical cleaning/
disinfecting and sub-chapter:
■
Chapter Mechanical neutral
or mild alkaline cleaning
and thermal disinfecting
13