Dental compressor with suction unit DUO, DUO 2, DUO 2V (Fig.4, Fig.5)
After unpacking the product from a packaging place it with it base onto the floor in a room, free it from
packaging materials and remove fixation parts (X,Y) - detail A. Place the wall-mounted stopper (28) - 2 pcs
onto the compressor housing in the rear top part of the housing and put the housing onto a required place.
The stoppers provide a sufficient distance of the housing from a wall for thorough ventilation. Open the door
on the housing using the attached key. If necessary, disassemble the doors by pulling the pin of door hinge
(32). Connect the equipment via ready made distributions in a floor according to installation plan or via the
holes on the rear side of the housing (Chapter
from the set (Chapter 9.2). Remove the connecting reinforcement (27) in the front part of the housing.
Thread the pressure hose via a hole in housing and connect it to an appliance in a suitable way (Chapter
9.3). Grasp the compressor at its handle and put it into the casing using built-in wheels (33) so that the front
part of the base was ca 20mm from the connecting reinforcement (27). Fix the hose (40) of a manometer
(30) in a hose into the fast-on coupling on a compressor, put the connecting reinforcement (27) back and
connect the pressure hose to a compressor. Insert the electric power cord (39) of a compressor into a socket
(34) on a housing and insert the free cord to a clip (38). By slight rotation of rectification screws (6) set the
correct position of door against the casing frame. When closing the door the pin (20) on the door must easily
snap in the opening in the casing frame. Close the housing doors and duly lock the lock (26). Connect the
mains plug into the mains socket.
It is not allowed to leave a key in a lock! It is necessary to keep it back from uninstructed persons!
Dental compressor with suction unit DUO/M, DUO 2/M, DUO 2V/M (Fig.4, Fig.5)
After unpacking the product from a packaging place it with it base onto the floor in a room, free it from
packaging materials and remove fixation parts (X, Y) - detail A. Place the compressor into housing similarly
as in the previous paragraph. Prior to placing the compressor into housing, protrude house for condensate
drain via hole in housing (24) and connect it to a bottle (22). The magnetic holder (31) with a vessel (22), for
entrapping condensate from a dryer may be fixed at the sides of housing or from the front on its doors. When
fixing the holder with a vessel at the housing side it is necessary to consider a space of at least 11 cm
between the housing and furniture. Distance smaller than the specified one may cause problem with
handling of the vessel.
The vessel must always be installed so that the lower section is near the floor; any other
installation may damage the dryer!
Noise suppressor for casing DUO, DUO2 (Fig.6)
Fix the suppressor (1) to the base of aspirator
using 4 screws M5 and washers. Interconnect
the output (2) from the
suppressor using the delivered hoses with ø30 –
400mm. Interconnect the output (3) from a
suppressor with the original output hose from
the aspirator. Lace the hoses via clips (4).
NP-DUO-4_11-2015-MD
9. 1). Connect the exhauster control cord of 24 V AC/DC
aspirator with a
Fig.6
- 32 -