This function makes using the device very simply
when the stored standards have a relative high colour
differ-ence.
6.2

Geometry

Under the menu "Geometry", the probe head version
can be selected. Also the "Stray light suppression"
function can be activated.
To manually select the
probe head adapter, go to
"Settings" menu and choose
"Geometry".
In the next window select
"Probe head" and press
"Enter".
It opens another selection
window. Choose the desired
selection with the "Enter"
key. The choice "Auto"
automatically detects the
connected probe head adapter
(factory setting).
The "Stray light suppression" measures any ambient
light that might be entering the probe head. This "stray"
light can affect the reproducibility and so this option
should be used if the opening of the sensor is not fully
covered by the sample or especially when measuring
transparent materials such as glass or translucent
materials. To activate the "Stray light suppression",
proceed as follows:
6.3

Illumination

The standard illuminant for calculating the colour value
scan be selected or changed.
ColorLite GmbH | sph870/sph900
Please choose "Geometry" from
menu "Setting". Turn on the "Stray
light suppression" by selecting the
appropriate menu item and confirm
with „Enter".
The active standard illuminant is
displayed in the menu position
"Illumination". The default setting
is the Daylight illumination "D65".
Geometry
71