5.1.2 Calibration with BAM- and
Working-Standard
If the option "Working-Standard" is activated the
calibration menu of the sph xs1 will show a second
menu option - "Working-Standard" - as well as the
"BAM-Standard".
It is important that the two white standards are
measured in the correct order.
Measure the BAM standard and then directly the
working standard in one procedure.
Select "BAM-standard" by pressing the "ENTER"
key. Follow the displayed instructions:
First calibrate with the BAM-
Standard.
Second you will be prompted
to calibrate with the Working-
Standard.
Again, the calibration is monitored. If a fault is detected,
an error message is displayed and you will be asked to
repeat the calibration.
Repeat the calibration with the BAM- and Work-ing-
Standard every 3 to 6 months according to use.
For the normal day to day calibration use only the
Working-Standard.
5.1.3 Calibration with the d/8° BAM-Standard
The BAM-Standard is traceable to the normal of the PTB
(Physikalisch-Technische Bundesanstalt) in Braunschweig
and is delivered with a certificate from the BAM (Federal
Institute for Materials Research and Testing) [see also 6.7
Colour systems].
ColorLite GmbH | sph870/sph900
The calibration is performed
in the same way as described
above. Please follow the displayed
instructions.
Calibration
59