Liability Statement - Cooper Surgical LEEP PRECISION LP-10-120 Bedienungsanleitung

Leep precision integrated system
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 41
Recommended Separation Distance
Recommended Separation Distance between portable and mobile RF communications equipment and
the CooperSurgical LEEP PRECISION Integrated System .
The CooperSurgical LEEP PRECISION Integrated Systems are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled . The customer or the user of the CooperSurgical
LEEP PRECISION Integrated Systems can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the CooperSurgical
LEEP PRECISION Integrated Systems as recommended below, according to the maximum output power of the
communications equipment .
Rated maximum
output power of
transmitter
in watts
0 .01
0 .1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer .
For the CooperSurgical LEEP PRECISION Integrated System:
V
= 3 Vrms
1
E
= 3 V/m
1
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
NOTE 2:
These guidelines may not apply in all situations . Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Section 17 Liability Statement
CooperSurgical guarantees the safety, reliability, and performance of the LEEP PRECISION Integrated System
only if the installation, recalibrations, and repairs are performed by personnel authorized by CooperSurgical
and if it is used in compliance with given instructions in an area that meets all applicable IEC requirements .
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
[ ]
3 .5
d=
P
V
1
0 .1167
0 .3689
1 .1667
3 .6894
11 .667
(Meters) [Notes 1 & 2]
80 MHz to 800 MHz
[ ]
3 .5
d=
P
E
1
0 .1167
0 .3689
1 .1667
3 .6894
11 .667
38
800 MHz to 2 .5 GHz
[ ]
7
d=
P
E
1
0 .2333
0 .7379
2 .3333
7 .3789
23 .333