Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance Procedures; Cleaning Procedures; Cautions - Cooper Surgical LEEP PRECISION LP-10-120 Bedienungsanleitung

Leep precision integrated system
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 41
4 . At the completion of each procedure activate the system to ensure safe particle containment . Using gloves
and a mask, remove the Pre-Filter, the Reducer, and the used section of Suction Tubing and
discard into an infectious waste receptacle (see Cautions) . The CooperSurgical Smoke Evacuator should
be stored with a new Pre-Filter and Reducer in place on the ULPA filter.
NOTE: The expected life of the ULPA filter is 3–6 months depending on usage. It must be discarded into
an infectious waste receptacle if plume odor is detected or the suction is diminished .

14.5 Maintenance Procedures

No maintenance is required during normal operation other than making sure that ample space is maintained
around the unit to allow for free air flow and adequate cooling.

14.6 Cleaning Procedures

The outside of the unit may be cleaned as necessary with a soft cloth dampened (not wet) with isopropyl
alcohol . Do not allow liquid to get inside of the unit .

14.7 Cautions

U .S . Federal law restricts this device to sale by or on the order of a physician .
This device produces a strong vacuum force and care should be exercised to ensure that the suction
control and the position of the inlet end of the suction tubing are adjusted properly to prevent injury to
the patient or inadvertent damage to surgical materials .
The materials in the plume removed by this device are potentially hazardous . Handle according to
29 CFR 1910 .1030 and OSHA 3127 .1992 (Occupational Exposure to Bloodborne Pathogens) guidelines .
To prevent a fire or explosion hazard, do not use the system in the presence of flammable or potentially
flammable materials.
Do not allow fluid to be pulled into the system.
To prevent premature failure of the ULPA filter cylinder, do not operate this device without a disposable
Pre-Filter in place .
34

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Leep precision lp-10-220

Inhaltsverzeichnis