2.7.2.1 Patient Return Electrode Use
There are two varieties of Patient Return Electrodes that can be used with the
LEEP PRECISION Integrated System:
1
Patient Return Electrode
With Connector
The Cable is built in
(
)
1
: If the Patient Return Electrode with the built-in cable is purchased, the Patient Return
Electrode Cable is plugged directly into socket
(
)
2
: If the Patient Return Electrode and a separate Reusable Cable are purchased, the
Patient Return Electrode is connected to the Reusable Cable, which is then plugged
into socket
.
8
Section 3 LEEP PRECISION Generator Features
Electronic power level control
Digitally derived waveforms
Low inherent leakage rate
Patient return monitor disables power automatically if Patient Return Electrode is loose
Regulated system performance traceable to NIST Standards
Automatic smoke evacuator signal function sends signal to detector for automatic smoke evacuator actuation
when generator is activated
Self-monitors all detectors
Classification
Model
LEEP PRECISION Generator
Do not get fluid into the LEEP PRECISION Generator. Should any liquid or solid object fall into the unit, unplug
the unit and call Technical Support .
The LEEP PRECISION Generator is suitable for intermittent operation with a two-minute ON and six-minute
OFF duty cycle .
The LEEP PRECISION Generator is classified as normal equipment (IPX0) according to protection against
ingress of water .
2
Patient Return Electrode
Without Connector
Reusable Cable is purchased
separately from CooperSurgical
.
8
Safety Class
I
11
Type