Section 14 LEEP PRECISION Smoke Evacuator

14.1 System Description

The CooperSurgical LEEP PRECISION Smoke Evacuator (three-stage air filtration system) is used to remove
airborne particulate plumes produced during office and surgical procedures and has the following features:
Low noise level .
Triple filtration of air provides efficiency level for 0.014 microns rated at 99.999 percent. This includes
a Pre-Filter, a charcoal filter for odor removal, and a final safety filter placed after the charcoal filter.
High air flow for plume collection.
Attaches to the CooperSurgical LEEP PRECISION Integrated System .

14.2 Initial System Setup

The system arrives in one carton that contains the Vacuum Unit . The
disposable accessories necessary to conduct procedures may be purchased
separately from CooperSurgical .

14.3 ULPA Filter Installation

Tilt the LEEP PRECISION Smoke Evacuator forward and insert the large
ULPA Filter cylinder with the Air Flow arrow pointing downward (see Photo A) .

14.4 Operating Instructions

Photo B
For procedures requiring close proximity plume removal (i.e., vaginal speculum)
Assemble the ⅜-inch Reducer (Catalog number 6083) onto the port on the
disposable Pre-Filter top with a slight twisting motion . Attach one end of an
appropriate length of ⅜-inch inner diameter (ID) Evacuation Tubing (Catalog
number 6084) to the Reducer Connector and direct the other end to the patient
and any appropriate device being used, such as a vaginal speculum equipped
with a Smoke Evacuation Adapter (see Photo C) .
For procedures requiring open area plume removal (i.e., external lesions)
1 . Assemble the sterile Disposable Evacuation Tubing (Catalog number 6084)
directly into the top of the Pre-Filter . Position the opposite end over the site
to be treated .
2 . Make sure the LEEP PRECISION Smoke Evacuator is plugged into the LEEP PRECISION Integration Unit
(refer to Section 2) .
3 . Push the POWER switch on the LEEP PRECISION Integration Unit to the ON position to start the LEEP
PRECISION Smoke Evacuator . The LEEP PRECISION Smoke Evacuator will activate upon use of the Foot
Switch or Active Electrode (Pencil) .
14.4.1 Installing the Pre-Filter
Prepare the unit for use by inserting a clean, disposable Pre-Filter (Single
Patient Use) onto the ULPA Filter cylinder (see Photo B) .
Be sure this device is firmly seated.
14.4.2 Connecting the Tubing
There are two tubing hookup options:
1 . For procedures requiring close-proximity plume removal
(i .e ., vaginal speculum)
2 . For procedures requiring open-area plume removal
(i .e ., external lesions)
33
Photo A
Photo C