8.2.1 Changing the prosthesis setting using the cockpit app
Product
MyModes
Functions
Settings
Status
General
Manage components
Imprint/Info
8.2.2 Overview of adjustment parameters in basic mode
The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait
cycle. These parameters act as basic settings for automatically adjusting the damping behaviour
to the current motion situation (e.g. ramps, slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further informa
tion on the stance function (see Page 66). Further information on the sitting function (see
Page 67).
The following parameters can be modified:
Parameter
Resistance
Stance function
Sitting function
70
1) Once the component is connected and in the desired mode,
tap the symbol in the main menu.
→ The navigation menu opens.
2) Tap the menu entry "Settings".
change
→ A list appears with the parameters for the currently
2. Golf
3) Change the setting of the desired parameter by tapping the
"<", ">" symbols.
INFORMATION: The prosthetist's setting is marked
and, after the setting has been changed, can be
restored by tapping the "Standard" button.
Adjustment
software
range
120 to 180
selected mode.
Setting
Meaning
range, app
+/- 10 of the
Resistance against flexion e.g. when
configured
walking down stairs or when sitting
value
down
0/Off - deactiv
Activation/deactivation of the stance
ated
function. This function needs to be
1/On - activ
enabled in the adjustment software.
ated
0/Off - deactiv
Activation/deactivation of the sitting
ated
function. This function needs to be
1/On - activ
enabled in the adjustment software.
ated