Herunterladen Inhalt Inhalt Diese Seite drucken

Instructing The Patient In The Use Of The Product - Ottobock 8S710-bebionicSkinSilicone Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Removal with re-use of the prosthetic glove
Removal without re-use of the prosthetic glove
6 Use

6.1 Instructing the patient in the use of the product

INFORMATION
Inform the patient about all the important points relevant to them that are contained in these
instructions for use.
Instruct the patient in the correct use of the product in accordance with the Instructions for
use (patient) included in the scope of delivery.
Hand over the Instructions for use (patient) to the patient when the briefing has been com­
pleted.
► NOTICE! Do not use 640F18 donning spray for application
and removal.
► NOTICE! Do not fold over the sleeve end for application
and removal.
► NOTICE! Avoid all excessive force on the fingers (espe­
cially the thumb) during the removal process.
Carefully pull the prosthetic glove off the prosthetic hand.
► NOTICE! Avoid damaging the prosthetic hand during cut­
ting.
Cut open the prosthetic glove on the inside.
► NOTICE! Avoid all excessive force on the fingers (espe­
cially the thumb) during the removal process.
Pull the sleeve apart and carefully pull off the prosthetic glove.