Detailed EMC Information

The equipment has been tested and found to
comply with the EMC limits for Medical Devices
per EN 55011 Class B and EN 60601-1-2. These
limits are designed to provide reasonable
protection against harmful interference in a
typical medical installation. Operation is subject
to the following two conditions: 1) this VP4726
or VP4726F Monitor may not cause harmful
interference, and (2) this VP4726 or VP4726F
Monitor must accept any interference received,
including interference that may cause undesired
operation.
The Equipment generates and can radiate radio
frequency energy and, if not installed and used
in accordance with the instructions, may cause
harmful interference to other devices in the
vicinity. However, there is no guarantee that
interference will not occur in a particular
installation.
If this equipment does cause harmful interference
with other devices, which can be determined by
turning the equipment off and on, the user is
encouraged to try to correct the interference by
one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the
equipment.
Connect the equipment to an outlet on a
circuit different from that to which the
other device(s) is connected.
Consult the manufacturer or a field service
technician for assistance.

FCC WARNING :

This equipment generates or uses radio
frequency energy. Changes or modifications to
this equipment may cause harmful interference
unless the modifications are expressly approved
in the instruction manual. The user could lose
authority to operate this equipment if an
unauthorized change or modification is made.
NOTES:
1.
The EMC tables and other guidelines
that are included in the Instruction
Manual provide information to the
customer or user that is essential in
determining the suitability of the
Equipment or System for the Electro-
magnetic Environment of use, and in
managing the Electromagnetic
Environment of use to permit the
Equipment or System to perform its
intended use without disturbing other
Equipment and Systems or non-medical
electrical equipment.
2.
Medical Electrical Equipment needs
special precautions regarding EMC and
needs to be installed and put into service
according to the EMC information
provided in the Accompanying Documents.
E-41
E

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis