15.4

Relative centrifugal force (RCF) and centrifuging radius (RAD)

Relative centrifugal force (RCF) depends on the centrifugal radius (RAD). Before RCF is set, the centrifugal
radius has to be set.

Press the
key as often as is required until the parameters RAD and RCF are displayed and the value of
RCF
the parameter, RAD is displayed in parentheses,  e.g. RAD = 146

Use the adjusting knob
By changing the centrifuging radius, the value adjusts automatically to the RCF.

key again. The value of the parameter, RCF is displayed in  parentheses, e.g. RAD = 146
Press the
RCF
RCF = 3695.

Press the
key to save the set RCF value. See the "Inputting or changing programs" chapter.
PROG

Press the
or
RCF
START
Only by storing (STO) the set RCF value will the RPM value resulting from that be accepted.
15.5

Start-up and run-down parameters

The set start-up and run-down parameters are displayed.
x: 1-9 = start-up stage, t = start-up time
y: 1-9 = braking stage, 0 = unbraked run-down, t = run-down time
y
x
15.5.1

Start-up stage and start-up time

It is only possible to set start-up times if these have been activated. See the chapter "Activating or deactivating
start-up and run-down times".

Press the
key. The parameter
= start-up stage,
t
= start-up time
Press the
key to switch between the start-up stage and start-up time.
TIME

Set the desired stage or time with the rotary knob .

If necessary, press the

To apply the setting to the display, either press the
the centrifugation data are displayed.
15.5.2

Braking stage and run-down time

No B-brake stages can be set for this device. The activation of the B-brake stages in the "Settings" menu is
not possible (parameter B-Ramp = off). B-brake stages are similar to an exponential braking curve.
It is only possible to set run-down times if these have been activated. See the chapter "Activating or
deactivating start-up and run-down times".

Press the
key until the parameter
= braking stage,
= run-down time
t
Press the
key to switch between the braking stage and run-down time.
TIME

Set the desired stage or time with the rotary knob .

If necessary, press the

To apply the setting to the display, either press the
the centrifugation data are displayed.
15.5.3

Brake switch-off speed


Press the
key as often as necessary until the parameter N Brake is shown.

Use the adjusting knob

Press the
or
START
62/155
to set the centrifuging radius you want.
keys to apply the setting to the display.
or
t
is displayed.
key to set the next parameter.
START
or
t
is displayed.
key to set the next parameter.
START
to set the value you want.
keys to apply the setting to the display.
RCF = 3695. The LED is lit in the key.
key or press the
key as often as is required until
key or press the
key as often as is required until