Herunterladen Inhalt Inhalt Diese Seite drucken

Cooling (Only In Centrifuges With Cooling); Standby-Cooling; Precooling Of The Rotor; Time-Controlled Cooling - Hettich ROTANTA 460 Robotic Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

31

Cooling (only in centrifuges with cooling)

The temperature set-point can be adjusted from -20°C to +40°C / -4°F to +104°F. The lowest obtainable temperature
is dependent on the rotor (see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories").
31.1

Standby-cooling

When the rotor is at a standstill, with the lid and hatch closed, the centrifuging chamber is cooled to the preselected
temperature if this is lower than 20°C / 68°F.
During the standby cooling, the preset temperature is displayed.
31.2

Precooling of the rotor

For quick precooling of the unloaded rotor and the accessories, it is a good idea to do a centrifugation run with
the continuous operation setting and a speed of about 20% of the maximum rotor speed.
A centrifugation run for pre-cooling the rotor can be carried out via the operating unit (key position "0") or via
the interface (key position "LOCK 2"). The functions of the various key positions are described in the "Key-
operated switch" chapter.
The centrifugation run for pre-cooling the rotor is done automatically with the program PREC (PRECOOLING).
A centrifugation run for pre-cooling the rotor cannot be carried out if program linkages are used.

Press the
key. The LED in the

After the time expires or if the centrifugation run is aborted via the interface or by pressing the
the run-down is done with the selected run-down parameter. The run-down parameter is displayed e.g.
right-hand LED in the
OPEN / STOP
the right LED in the
OPEN / STOP
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber and the
remaining or elapsed time are displayed.
31.3

Time-controlled cooling

If necessary it can be set so that cooling begins after a certain amount of time once centrifugation run has been
started.
The delay time can be set in 1-second increments from 15 to 900 seconds. If no time delay is required, set to "0". No
delay time is set ex works.
When the rotor is at a standstill, the delay time can be set as follows:
This setting can only be made via the operating unit (key position "0" or "LOCK 2"). The functions of the
various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Settings is displayed.
PROG

Press the
key. The SOUND / BELL = off/on is displayed.
START

Press the
key as often as is needed until the delay time is displayed, e.g. Cool acc time = 0.
PROG

Use the adjusting knob
0 = no delay time

Press the
key to save the setting.
START
Briefly, Store Settings ... is displayed as confirmation, followed by -> Settings.

Press the
key once to exit the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
key blinks until the rotor has been imported, it is subsequently lit.
key is lit. Once the rotor has come to a standstill, the LED in the
key switch off and the left LED in the
to set the value you want.
key lights up.
OPEN / STOP
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
key twice to exit the
OPEN / STOP
EN
key,
OPEN / STOP
9. The
key and
73/155

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis