Herunterladen Inhalt Inhalt Diese Seite drucken

Entering Centrifugation Parameters; Direct Input Of The Centrifugation Parameters; Speed (Rpm); Relative Centrifugal Force (Rcf) And Centrifugal Radius (Rad) - Hettich MIKRO 185 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für MIKRO 185:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
16

Entering centrifugation parameters

16.1

Direct input of the centrifugation parameters

The speed (RPM), the relative centrifugal force (RCF), the centrifuging radius (RAD) and the runtime can be input
directly with the keys
The set centrifugation parameters are only stored after starting the centrifugation run.
16.1.1

Speed (RPM)

 Press the key to activate the RPM display
RCF
(RPM) as needed.
 Set the desired value with the keys.
RPM/RCF
16.1.2

Relative centrifugal force (RCF) and centrifugal radius (RAD)

 Press the key to activate the RCF display
RCF
( RCF ) as needed.
 Set the desired RCF value with the keys.
RPM/RCF
 Set the desired centrifuging radius with
t
the keys as needed.
16.1.3

Runtime

Up to 1 minute, the runtime can be set in steps of 1 second, and starting from 1 minute, it can only be set in
steps of 1 minute.
In order to set the continuous run, the parameters t/min and t/sec must be set to zero. In the time display
(t/min:s), "--:--" appears.
Example:
>RCF<
RPM
t/min:s
 Set the desired value with the keys.
t
54/142
without previously having to press the
Example:
Example:
Example:
SELECT
key.
>RCF<
RPM
t/min:s
>RCF<
RPM
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
>RCF<
RPM
t/min:s
>RCF<
RPM
t/min:s

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis