Do not advance the guidewire or catheter if unusual resistance is
encountered.
Do not insert or withdraw the guidewire forcibly from any
component. The wire may break or unravel. If the guidewire
becomes damaged, the introducer needle or sheath/dilator and
guidewire must be removed together.
Federal Law (USA) restricts this device to sale by or on the order of a
physician.
This catheter is for Single Use Only.
Do not re-sterilize the catheter or
accessories by any method.
Re-Use may lead to infection or illness/injury.
The manufacturer shall not be liable for any damages caused by
reuse or re-sterilization of this catheter or accessories.
Contents sterile and non-pyrogenic in unopened, undamaged
package.
STERILIZED BY ETHYLENE OXIDE
Do not use catheter or accessories if package is opened or damaged.
Do not use catheter or accessories if any sign of product damage is
visible.
DO NOT use high-pressure injectors for contrast medium studies.
Excessive pressures may damage catheter.
CATHETER PRECAUTIONS:
Small syringes will generate excessive pressure and may damage
the catheter. Ten (10)cc or larger syringes are recommended.
Do not use sharp instruments near the extension lines or catheter
lumen.
Do not use scissors to remove dressing.
Catheter will be damaged if clamps other than what is provided with
this kit are used.
Clamping of the tubing repeatedly in the same location will weaken
tubing. Avoid clamping near the luer(s) and hub of the catheter.
Examine catheter lumen and extension(s) before and after each
infusion for damage.
To prevent accidents, assure the security of all caps and
connections prior to and between treatments.
Use only Luer Lock (threaded) Connectors with this catheter.
Repeated over tightening of luer lock connections, syringes, and
caps will reduce connector life and could lead to potential connector
failure.
Confirm catheter tip position by x-ray prior to use. Monitor tip
placement routinely per institution policy.
INSERTION SITES:
The basilic, median cubital, or cephalic vein may be catheterized.
The basilic vein is the preferred site.