Lithium cell replacement
Lithium cell replacement
Lithium cell replacement
Lithium cell replacement
Lithium cell replacement
Clock problems are in most cases
caused by a weak lithium cell. To
replace the cell, open the unit as
described under chapter "fault finding
procedures". Use a small screw driver
to lift the cell from its socket. Remove
the cell by sliding it away. Replace by
3 Volts cell type CR2025 only.
Software up date
Software up date
Software up date
Software up date
Software up date
To up date the software of the MCX, you
must replace one or more EPROMS
(Erasable Programmable Read Only
Memory) in accordance to the instructions
issued. The EPROM's are identified and
located as indicated. Versions are
indicated by date labels on the EPROM.
Read under
chapter "fault finding procedures" how to
open and close the unit. Actual installed
EPROM version dates are displayed when
the unit is switched on.
Updates are required to adapt the
MCX to receive new options or for best
performance. Ask Unomat Instrumenten
BV or your local sales point for latest
updates.