Recalibration
Recalibration
Recalibration
Recalibration
Recalibration
NOTES:
1.
Calibration standards should have
a valid calibration certification
and should be at least 3 times
more accurate as the published
accuracies for the MCX for
each range.
2.
As derived from electrical
parameters, temperature sensor
ranges do not require any
individual recalibration.
3.
Be sure that you use the test
leads, as standard supplied with
the MCX or other low e.m.f.
test leads.
4.
The MCX should be powered
from the on board batteries
during recalibration.
5.
Recalibration must be carried out
at reference ambient temperature
and humidity. MCX must be
exposed to this condition for at
least 2 hours before the actual
recalibration starts.
6.
Recalibrate mV ranges before
starting recalibration of the Cold
Junction Compensation system.
7.
When accessing the calibration
menu the temperature unit is set
to °C and the temperature scale
is set to IPTS 68.
8.
Calibration is recommended
between 12 and 14 months.