Recalibration of the MCX

Recalibration of the MCX
Recalibration of the MCX
Recalibration of the MCX
Recalibration of the MCX
Leave the cursor at NONE in the INPUT
and OUTPUT menu and press the # and
EXE keys simultaneously to open the
settings menu. Select CALIBRATION
and press EXE. Fill out the access code
and press EXE to open the calibration
menu.
The access code is fixed at 9410
and works as an barrier to prevent
accidental access to MCX
calibration ranges.
The calibration menu shows all relevant
ranges to calibrate. Each range can
be recalibrated independent of any
other range. Only the cold junction
compensationsystem is influenced
by millivolts recalibration.
Select the ranges to be calibrated and
press EXE. Follow the instructions on
the next screens. Fill out the calibration
readings and press EXE to confirm or
to return to the calibration menu.
Select an other range to continue
recalibration or press CE to leave
the calibration menu.