2. Insert a test strip into the test strip guide on the
1
monitor
Remove a fresh test strip from its foil pouch. Insert it into the
test strip guide so that the clear end with the vertical contact
bars goes in first. On the blue end of the strip you should be able
to read the word "INRatio®" appearing from left to right. Make
sure the sample well in the strip lines up with the green light on
the test strip guide.
3. Match the code on the monitor display with the
strip code on the test strip pouch or strip container
If the codes match, simply press the
flashing on the display and proceed to Step 4.
If the codes do not match, use the and buttons to change
the first digit of the code which will then be flashing.
Press the
next digit. Change each digit using the and buttons, then
press the
digit.
To scroll more quickly through letters and numbers while
making changes, hold the and buttons down.
After setting the fifth digit, "OK?" will begin flashing on the
display. If the code is correct, press the
code in the monitor.
Or, if you need to correct the code, press the and buttons to
go back to the first digit and repeat step 3 until the correct code
is on the monitor display.
button while "OK? "is
button to accept the first digit and move to the
button to accept each digit and move to the next
button to store the
CAUTION: Always match the code on the monitor display with the strip
code on the test strip pouch/container. Failure to do so may cause
inaccurate results.
NOTE: If you entered an invalid strip code, the CAUTION
symbol (
) will flash underneath the code. Re-enter the
correct code beginning with the first digit, which is now
flashing.
4. Wait for the monitor to warm up
The monitor will count down for a few seconds and display
the Temperature symbol ( ) as it warms up for the test.
The monitor display will prompt you to apply a blood sample
when it's ready to perform a test.
5. Get a fingerstick sample
See Instructions for Collecting a Fingerstick Sample (Section 9) for getting the correct blood
sample to perform a test.
6. Apply the blood sample to the sample well on the
test strip
Make sure you apply the sample directly over the green light.
DO NOT move the monitor or your finger as you apply the
sample.
If you applied a blood sample before the monitor was ready, the
CAUTION symbol (
test strip. DO NOT add more blood to the test strip. Go back to
step 2 and re-test with a fresh test strip.
) will flash on the display. Remove the
EN