Display Screen
2
Ambient temperature is too
hot (above the operating
range) for the monitor to
work properly.
The monitor, strips and/or
ambient temperature is too
cold (below the operating
range) to continue with the
test.
The monitor, strips and/or
ambient temperature is too
hot (above the operating
range) to continue with the
test.
You applied a blood sample
to the test strip before the
monitor warm-up was
complete, or you inserted a
used test strip.
You did not apply enough
blood to fill the sample well
area.
What it Means
Action to Take
Move the monitor to a cooler
location and try again in a
few minutes.
Make sure the monitor and
strips are within the
operating range of the
monitor. Move the
monitor and strips to a
warmer location, then repeat
the test in a few minutes
with a fresh strip.
Make sure the monitor and
strips are within the
operating range of the
monitor. Move the monitor
and strips to a cooler
location, then repeat the
test in a few minutes with a
fresh strip.
Repeat the test with a fresh
test strip. Make sure to wait
until you are prompted to
apply a blood sample.
Repeat the test with a fresh
test strip. DO NOT add more
blood to the test strip.
Display Screen
What it Means
The strip code is invalid.
The QC1 test result is below
the control range of the
monitor.
The QC1 test result is above
the control range of the
monitor.
The QC2 test result is below
the control range of the
monitor.
The QC2 test result is above
the control range of the
monitor.
There is an error in
calculating Prothrombin
Time (PT) for this test. There
may be a problem with the
monitor and strips.
EN
Action to Take
2
Re-enter the correct code
beginning with the first
digit.
Repeat the test with a fresh
test strip. If this message
continues, call HemoSense
Technical Support for
assistance.
Repeat the test with a fresh
test strip. If this message
continues, call HemoSense
Technical Support for
assistance.
Repeat the test with a fresh
test strip. If this message
continues, call HemoSense
Technical Support for
assistance.
Repeat the test with a fresh
test strip. If this message
continues, call HemoSense
Technical Support for
assistance.
Repeat the test with a fresh
test Strip. If this message
continues, call HemoSense
Technical Support for
assistance.